Laserfiche WebLink
- - ----Pnvironmental <br /> Measurement Procedure <br /> 1. Rinse probe with DS/DI water. <br /> 2. Place probe in the solution to be measured. <br /> 3. Measure and record temperature. <br /> 4. Set MODE control knob to "X100." <br /> If the reading on the meter display is below 50 on the 0-500 micromho range, <br /> switch MODE control knob to "X10." <br /> If the reading on the display is still below 50 micromhos, switch the MODE <br /> control knob to "Xl." <br /> NOTE: The "micromho" designation on the meter display is a shorthand form for <br /> "micromhos/centimeter (µmhos/cm)." <br /> 5. Record conductivity value in µmhos/cm. Multiply the reading on the meter display <br /> by the appropriate scale factor. <br /> Example: <br /> Meter Display Reading = 245 <br /> • <br /> CO2129 <br />