Laserfiche WebLink
_ _ = v,=Pnvirorunental <br /> MODE Scale = X10 <br />_. Conductivity Value = 2,450 µmhos/cm <br />- 6. Correct the meter conductivity reading to a temperature of 25.0°C by finding the <br /> sample solution's temperature on Table D.3 and the corresponding multiplication <br /> - factor. <br /> Example: Sample temperature = 20.1°C <br /> Multiplication factor = 1.103 (from Table D.3) <br /> Temperature corrected conductivity = 2450 µmhos/cm x 1.103 <br /> = 2702.4 µmhos/cm <br /> 7. Multiply the temperature corrected conductivity value by the probe Correction Factor, <br /> calculated in Step No. 8 of the Probe Calibration Procedure, to get the solution <br /> conductivity value. <br /> Example: Solution conductivity = 2702.4 µmhos/cm x probe Correction Factor <br /> Probe Test <br /> When measuring on the X100 or X10 scales, depress the CELL TEST button. The meter <br /> reading should fall no more than 2 percent; if it is greater, the probe is fouled and the <br /> measurement is in error. Clean the probe and remeasure. NOTE: The CELL TEST does not <br /> function on the X1 MODE scale. <br /> • <br /> CO2130 <br />