Laserfiche WebLink
' Draft: April S, 1999 Page C-13 <br />C.6.2 Conductivity Meter <br />The conductivity meter must be calibrated each day before taking field measurements. <br />' Record time, temperatwe, and instrument response in the field notebook. Calibration is done by <br />noting the response of the meter to several standazd conductivity solutions which bracket the values <br />' expected to be measured in the field. Standazds of 100, 1,000, and 10,000 should be adequate for <br />the samples expected. If the instrument has a calibration adjustment, set the response to match the <br />' standazds. Otherwise, simply record the instrument response to each standazd in the field notebook. <br /> <br /> <br /> <br /> <br />1 <br />1 <br />1 <br />1 <br /> <br />1 <br /> <br />1 <br /> <br />' Agapito Associates, Inc. <br />