The samples in this SDG were analyzed on an"as received"basis. <br /> SOP Reference <br /> Procedure for preparation,analysis and reporting of analytical data are controlled by GEL Laboratories LLC as <br /> Standard Operating Procedure(SOP).The data discussed in this narrative has been analyzed in accordance with <br /> GL-RAD-A-001 REV# 17. <br /> Calibration Information: <br /> Calibration Information <br /> All initial and continuing calibration requirements have been met. <br /> Standards Information <br /> Standard solutions for these analysis are NIST traceable or verified with a NIST traceable standard and used <br /> before the expiration dates. <br /> Sample Geometry <br /> All counting sources were prepared in the same geometry as the calibration standards. <br /> Ouality Control(00 Information: <br /> Blank Information <br /> The blank volume is representative of the sample volume in this batch. <br /> Designated QC <br /> The following sample was used for QC: 344616002(Sunday MW 1D). <br /> QC Information <br /> All of the QC samples meet the required acceptance limits with the following exceptions:The matrix spike and <br /> matrix spike duplicate, 1203055935(Sunday MW 1D)and 1203055936(Sunday MW 11)),do not meet the alpha <br /> relative percent difference requirement;however,they do meet the recovery requirement.The blank, <br /> 1203055933(MB),did not meet the beta detection limit due to keeping the blank volume consistent with the <br /> other sample aliquots. <br /> Technical Information: <br /> Holding Time <br /> All sample procedures for this sample set were performed within the required holding time. <br /> Sample Re-prep/Re-analysis <br /> None of the samples in this sample set required reprep or reanalysis. <br /> Chemical Recoveries <br /> All chemical recoveries meet the required acceptance limits for this sample set. <br /> Gross Alpha/Beta Preparation Information <br /> High hygroscopic salt content in evaporated samples can cause the sample mass to fluctuate due to moisture <br /> absorption.To minimize this interference,the salts are converted to oxides by heating the sample under a flame <br /> until a dull red color is obtained.The conversion to oxides stabilizes the sample weight and ensures that proper <br /> alpha/beta efficiencies are assigned for each sample.Volatile radioisotopes of carbon,hydrogen,technetium, <br /> polonium and cesium may be lost during sample heating. <br /> Recounts <br /> None of the samples in this batch were recounted. <br /> Page 79 of 94 <br />