Laserfiche WebLink
ASBESTOS and OTHER FIBERS by PCM:METHOD 7400,Issue 2,dated 15 August 1994-Page 4 of 15 <br /> collection area Ac=385 mmZ)for optimum accuracy.These variables are related to the action level <br /> (one-half the current standard),L(fibers/cc),of the fibrous aerosol being sampled by: <br /> t= <br /> Ac x E <br /> QxLx103 <br /> NOTE 1:The purpose of adjusting sampling times is to obtain optimum fiber loading on the filter. <br /> The collection efficiency does not appear to be a function of flow rate in the range of 0.5 <br /> to 16 L/min for asbestos fibers[7]. Relatively large diameter fibers(>3 µm) may exhibit <br /> significant aspiration loss and inlet deposition.A sampling rate of 1 to 4 L/min for 8 h is <br /> appropriate in atmospheres containing ca.0.1 fiber/cc in the absence of significant amounts <br /> of non-asbestos dust.Dusty atmospheres require smaller sample volumes(s400 Q to obtain <br /> countable samples.In such cases take short,consecutive samples and average the results <br /> over the total collection time.For documenting episodic exposures,use high flow rates(7 <br /> to 16 L/min)over shorter sampling times.In relatively clean atmospheres,where targeted <br /> fiber concentrations are much less than 0.1 fiber/cc,use larger sample volumes(3000 to <br /> 10000 Q to achieve quantifiable loadings.Take care,however,not to overload the filter with <br /> background dust.If>-50%of the filter surface is covered with particles,the filter may be too <br /> overloaded to count and will bias the measured fiber concentration. <br /> NOTE 2:OSHA regulations specify a minimum sampling volume of 48 L for an excursion <br /> measurement,and a maximum sampling rate of 2.5 L/min [3]. <br /> 5. At the end of sampling,replace top cover and end plugs. <br /> 6. Ship samples with conductive cowl attached in a rigid container with packing material to prevent <br /> jostling or damage. <br /> NOTE: Do not use untreated polystyrene foam in shipping container because electrostatic forces <br /> may cause fiber loss from sample filter. <br /> SAMPLE PREPARATION: <br /> NOTE 1:The object is to produce samples with a smooth(non-grainy) background in a medium with <br /> refractive index<- 1.46.This method collapses the filter for easier focusing and produces <br /> permanent(1-10 years)mounts which are useful for quality control and interlaboratory <br /> comparison.The aluminum"hot block"or similar flash vaporization techniques may be <br /> used outside the laboratory[2].Other mounting techniques meeting the above criteria <br /> may also be used (e.g.,the laboratory fume hood procedure for generating acetone vapor <br /> as described in Method 7400—revision of 5/15/85,or the non-permanent field mounting <br /> technique used in P&CAM 239[3,7-91).Unless the effective filtration area is known, <br /> determine the area and record the information referenced against the sample ID number <br /> [1,9-11]. <br /> NOTE 2:Excessive water in the acetone may slow the clearing of the filter,causing material to be <br /> washed off the surface of the filter.Also,filters that have been exposed to high humidities <br /> prior to clearing may have a grainy background. <br /> 7. Ensure that the glass slides and cover slips are free of dust and fibers. <br /> 8.Adjust the rheostat to heat the"hot block"to ca.70°C[2]. <br /> NOTE: If the"hot block"is not used in a fume hood,it must rest on a ceramic plate and be isolated <br /> from any surface susceptible to heat damage. <br /> 9. Mount a wedge cut from the sample filter on a clean glass slide. <br /> a. Cut wedges of ca.25%of the filter area with a curved-blade surgical steel knife using a rocking <br /> motion to prevent tearing.Place wedge,dust side up,on slide. <br /> NOTE: Static electricity will usually keep the wedge on the slide. <br /> b. Insert slide with wedge into the receiving slot at base of"hot block" Immediately place tip of <br /> a micropipet containing ca.250 µL acetone(use the minimum volume needed to consistently <br /> clear the filter sections) into the inlet port of the PTFE cap on top of the"hot block"and inject the <br /> NIOSH Manual of Analytical Methods(NMAM),Fourth Edition <br />