Laserfiche WebLink
fortified sample matrix covering the 3.5 Laboratory performance check 4.2 The water dip or negative peak that <br /> anions of interest. The fortification standards (LPC) - a solution of elutes near and can interfere with the <br /> procedure is described in Section 11 6 analytes prepared in the laboratory by fluoride peak can usually be eliminated <br /> adding appropriate volumes of the stock by the addition of the equivalent of 1 mL <br /> 2. Summary of Method standard solutions to reagent water of concentrated eluent (7.3 100X)to <br /> used to evaluate the performance of the 100 mL of each standard and sample. <br /> 2.1 A small volume of sample, typically instrument system with respect to a <br /> 2 to 3 mL, is introduced into an ion defined set of method criteria. 4.3 Method interferences may be <br /> chromatograph. The anions of interest caused by contaminants in the reagent <br /> are separated and measured, using a 3.6 Laboratory duplicates(LD) -two water, reagents,glassware, and other <br /> system comprised of a guard column, aliquots of the same sample that are sample processing apparatus that lead <br /> separator column, suppressor device, treated exactly the same throughout to discrete artifacts or elevated baseline <br /> and conductivity detector. laboratory analytical procedures. in ion chromatograms. <br /> Analyses of laboratory duplicates <br /> 2.2 The main differences between indicate precision associated with 4.4 Samples that contain particles <br /> Method A and B are the separator laboratory procedures but not the larger than 0.45 microns and reagent <br /> columns, guard columns and eluents. sample collection, preservation,or solutions that contain particles larger <br /> Sections 6 and 7 will elicit the differ- storage procedures. than 0.20 microns require filtration to <br /> ences. prevent damage to instrument columns <br /> 3.7 Field duplicates (FD) -two and flow systems. <br /> 2.3 In order to use this method for samples taken at the same time and <br /> solids an extraction procedure must be Placed under identical circumstances 4.5 Any anion that is not retained by <br /> performed (See 11.7). and treated exactly the same through- the column or only slightly retained will <br /> out field and laboratory procedures. elute in the area of fluoride and inter- <br /> 3. Definitions Analyses of field duplicates indicate the fere. Known coelution is caused by <br /> precision associated with sample carbonate and other small organic <br /> 3. collection, preservation and storage, as anions. At concentrations of fluoride <br /> co Stock standard solution a well as with laboratory procedures. above 1.5 mg/L this interference may <br /> concentrated solution containing a not be significant, however, it is the <br /> single certified standard that is a are u . Stock standard 3.8 Laboratory fortified sample matrix responsibility of the user to generate <br /> method a (LFM)-An aliquot of an environmental precision and accuracy information in <br /> solutions are used to prepare calibra sample to which known quantities of the each sample matrix. <br /> lion standards. method analytes are added in the <br /> laboratory. The LFM is analyzed exactly 4.6 The acetate anion elutes early <br /> solution <br /> Calibration standards (CAL) -a like a sample, and its purpose is to during the chromatographic run. The <br /> solution analytes prepared in the determine whether the sample matrix retention times of the anions also seem <br /> laboratoryy from stock standard solutions contributes bias to the analytical results. to differ when large amounts of acetate <br /> and diluted as needed and used to The background concentrations of the are present. Therefore,this method is <br /> calibrate the instrument response with analytes in the sample matrix must be not recommended for leachates of solid <br /> respect to analytic concentration. determined in a separate aliquot and samples when acetic acid is used for <br /> the measured values in the LFM pH adjustment. <br /> 3.3 Quality control sample (OCS) -a corrected for background concentra- <br /> solution containing known concentra- tions, 4.7 The quantitation of unretained <br /> tions of analytes, prepared by a peaks should be avoided, such as low <br /> laboratory other than the laboratory 3.9 Laboratory fortified blank (LFB) - molecular weight organic acids (for- <br /> performing the analysis. The analyzing An aliquot of reagent water to which mate, acetate, propionate, etc.)which <br /> laboratory uses this solution to demon- known quantities of the method are conductive and coelute with or near <br /> strate that it can obtain acceptable analytes are added in the laboratory. fluoride and would bias the fluoride <br /> identifications and measurements with The LFB is analyzed exactly like a quantitation in some drinking and most <br /> a method. sample, and its purpose is to determine waste waters. <br /> whether the methodology is in control, <br /> 3.4 Performance evaluation sample and whether the laboratory is capable 5. Safety <br /> (PE)- a solution of method analytes of making accurate and precise meas— <br /> distributed by the Quality Assurance urements at the required method de— <br /> Research Division (OARD), Environ 5.1 Normal, accepted laboratory safety <br /> practices should be followed during <br /> mental Monitoring Systems Laboratory tection limit. P 9 <br /> (EMSL-Cincinnati), USEPA, Cincinnati, reagent preparation and instrument <br /> Ohio,to multiple laboratories for 4. Interferences operation. No known carcinogenic <br /> analysis. A volume of the solution is materials are used in this method. <br /> added to a known volume of reagent 4.1 Interferences can be caused by <br /> water and analyzed with procedures substances with retention times that are 6. Apparatus and Materials <br /> used for samples. Results of analyses similar to and overlap those of the anion <br /> are used by the OARD to determine of interest. Large amounts of an anion 6.1 Balance-Analytical, capable of <br /> statistically the accuracy and precision can interfere with the peak resolution of accurately weighing to the nearest <br /> that can be expected when a method is an adjacent anion. Sample dilution 0.0001 g. <br /> performed by a competent analyst. and/or fortification can be used to solve <br /> Analyte true values are unknown to the most interference problems. <br /> analyst. <br /> 300.0-2 August 1991 <br />