Laserfiche WebLink
<br />ACZ Laboratories, Inc. August 10, 2007 <br />Quality Assurance Plan Version 12 ' <br />SOPAD018.08.07.12 Page 46 of 96 <br />• RPD for a matrix duplicate is evaluated only if the observed concentration is greater <br /> <br />than 10 times the MDL; otherwise each associated client sample must be reported ' <br />with the appropriate qualifier, regardless of RPD. <br />• In the absence of other contributing factors, a DUP failure for a solid or semi-solid ' <br />matrix is attributed to non-homogeneity of the sample, and each associated client <br />sample may be reported with the appropriate qualifier. <br /> <br />• For an aqueous matrix, if the DUP fails then all associated samples and the DUP <br />must be retested. If permitted by the instrument software the sample and DUP can <br />be reanalyzed at the end of the analysis in lieu of retesting all associated samples. ' <br />• For an aqueous matrix, if the MS/MSD RPD fails then the associated samples must <br />be reanalyzed. If permitted by the instrument software the sample and MS/MSD <br /> <br />can be reanalyzed at the end of the analysis in lieu of retesting all associated samples. , <br />• If applicable, evaluate the LCS/LCSD if the RPD fails for a matrix duplicate or spike <br />duplicate. Each associated client sample may be reported with the appropriate , <br />qualifier if the LCS/LCSD meets the criteria stated in 11.1.2.2. <br />• For a solid or semi-solid matrix, if both the LCSS and LCSSD recoveries pass but ' <br />the RPD fails, then acceptable precision may be demonstrated by a passing RPD for <br />the MS/MSD, and each associated client sample may be reported with the <br />appropriate qualifier. , <br />11.2 Instrument Specific Controls <br />All data must be associated with a passing instrument calibration and initial calibration verification. To the <br />extent possible, all data must be associated with passing continuing calibration verification. If the initial <br />calibration verification results (ICV/ICB) are outside of the acceptance criteria, then the source(s) of the <br />failure must be identified, corrective action(s) performed if necessary, and the instrument recalibrated before ' <br />proceeding with sample analysis. <br />If the continuing calibration verification results (CCV/CCB) do not meet the acceptance criteria then the ' <br />source(s) of the failure must be identified and corrective action(s) performed, including recalibration if <br />necessary, before continuing with sample analysis. If reanalysis of any sample(s) associated with failing <br />calibration verification is not possible then the associated data must be reported with the appropriate <br />qualification. ' <br />For instruments that permit the analysis of subsequent workgroups using the most recent calibration, two (2) <br />consecutive attempts of the opening CCV/CCB are allowed. If both attempts fail to produce acceptable results ' <br />then the source(s) of the failure must be identified and corrective action(s) performed, including recalibration if <br />necessary, before commencing sample analysis. <br />Unless stated otherwise by the test SOP, passing calibration verification must always bracket all batch quality <br />, <br />control samples, and results for additional instrument check standards, if applicable, must be within the <br />acceptance criteria stated in the SOP. However, when the acceptance criteria for a CCV or CCB are <br />exceeded (i.e. high bias) any associated client sample with a measured concentration less than the MDL may ' <br />be accepted and reported with the appropriate qualification. <br />2773 Downhill Drive 970-879-6590 ' <br />Steamboat Springs, CO 80487 www.acz.com