Laserfiche WebLink
IID. INORGANIC ANALYSIS WORKSHEET -- CN CALIBRATIONS // TECHLAW <br />BATCH. ~bI22~ <br />uq w ~ uie~ wo um nroet me Veiu:m iewvei arena mr Ire II.V anaror V V V sfanOdfO. <br />ICV TRUE Found %R Action Samples .~Hected <br />CCV <br />w <br />r <br />1. Were the corned number of standards and blanks used to calibrate the instrument] Yes No <br />2. Is the initial calibrations correlation coeKcient > 0.9957 Yes No (~_ ~ . Q ,qq q q T--II <br />13. Was the midrange calibration standartl or an ICV distilled? Yes No I <br />I(no, list aHeded <br />Actions: <br />PERCENT RECOVER <br /><70% 70-84% BS116% 176-130% >130% <br />Detected results R J V J R <br />NonEeteded Results R UJ V V V <br />1. If three standards and a blank were not used for initial calibration, ar the instrument was not calibrated daily and each time the instrument was set <br />up, qualify the data as rejected (R). <br />2. If the initial calibration comelalion coefficient was less than 0.995, quality sample results as estimated (J)/(UJ). <br />3. II a mid-range standard (or ICV) was not distilled, quality sample results as estimated (J)!(UJ). <br />Inorg98.xls <br />