Laserfiche WebLink
<br /> <br />5.2 Glass electrode. <br />5.3 Reference electrode-a calomel, silver-silver chloride or other reference electrode of <br />constant potential may be used. <br />NOTE 1: Combination electrodes incorporating both measuring and reference <br />functions are convenient to use and are available with solid, gel type filling materials that <br />require minimal maintenance. <br />5.4 Magnetic stirrer and Teflon-coated stirring bar. <br />5.5 Thermometer or temperature sensor For automatic compensation. <br />6. Reagents <br />6.1 Primary standard buffer salts are available from the National Bureau of Standards and <br />should be used in situations where extreme accuracy is necessary. <br />6.1.1 Preparation of reference solutions from these salts require some special precautions <br />and handling"' such as low conductivity dilution water, drying ovens, and carbon <br />dioxide free purge gas. These solutions should be replaced at least once each <br />month. <br />6.2 Secondary standard buffers may be prepared from NBS salts or purchased as a solution <br />from commercial vendors. Use of these commercially available solutions, that have been <br />validated by comparison to NBS standards, are recommended for routine use. <br />7. Calibration <br />7.1 Because of the wide variety of pH meters and accessories, detailed operating procedures <br />cannot be incorporated into this method. Each analyst must be acquainted with the <br />operation of each system and familiar with all instrument functions. Special attention to <br />care of the electrodes is recommended. <br />7.2 Each instrument/electrode system must be calibrated at a minimum of two points that <br />bracket the expected pH of the samples and are approximately three pH units or more <br />apart. <br />7.2.1 Various instrument designs may involve use of a "balance" or "standardize" dial <br />and/or a slope adjustment as outlined in the manufacturer's instructions. Repeat <br />adjustments on successive portions of the two buffer solutions as outlined in <br />procedure 8.2 until readings are within 0.05 pH units of the buffer solution value. <br />8. Procedure <br />8.1 Standardize the meter and electrode system as outlined in Section 7. <br />8.2 Place the sample or butler solution in a clean glass beaker using a sufficient volume to <br />cover the sensing elements of the electrodes and to give adequate clearance for the <br />magnetic stirring bar. <br />81.1 If field measurements are being made the electrodes may be immersed directly in <br />the sample stream to an adequate depth and moved in a manner to insure sufficient <br />sample movement across the electrode sensing element as indicated by drift free <br />(~ 0.1 pH) readings. <br />8.3 If the sample temperature differs by more than 2°C from the buffer solution the measured <br />pH values must be corrected. Instruments are equipped with automatic or manual <br />"'National Bureau of Standards Special Publication 260. <br />150.1-2 <br />