middle of the analytical range, a Continuing Calibration Verification (CCV) <br />standard with concentrations near the middle of the analytical range but <br />different than those in the ICV, and a readback of the highest calibration <br />standard. <br />These solutions provide verification that the cahbration equations aze <br />functioning properly throughout the analytical range of the instrument. During <br />sample analysis dilutions aze made for analytes found at concentrations above <br />the highest calbration standard. No results aze taken from extrapolations <br />beyond the highest standard. <br />6. All standazds and solutions are NTST traceable and were used within their <br />recommended shelf life. <br />7. The sample was prepazed and analyzed within the established hold times. <br />All in house quality control procedures were followed, as described below. <br />8. General quality control procedures. <br />^ A filter (method) blank and laboratory control sample were filtered and <br />preserved at the same time as the samples for the selenium analysis. A filter <br />(method) blank was filtered and preserved at the same time as the samples for <br />the thallium analysis. There were not more than 20 samples associated with <br />each fdtered blank and laboratory control sample. <br />^ The prepazation (method) blank associated with each digestion batch was <br />below the practical quantitation limit for each requested analyte. <br />^ The laboratory control sample associated with the selenium batch was within <br />the acceptance limits. This indicates complete digestion according to the <br />method. <br />^ All initial and continuing cahbration blanks associated with this analytical batch <br />were below the practical quantitation limits for the requested analytes. <br />^ All initial and continuing calibration verifications associated with this analytical <br />batch were within the acceptance criteria for the requested analyzes. This <br />indicates a valid calibration and stable instrument conditions. <br />^ The interference check samples associated with Method 200.8 were analyzed. <br />9. Matrix specific quality control procedures. <br />Per method requirements, matrix QC was performed for this analysis. Since a <br />sample from this order number was not the selected quality control (QC) sample, <br />matrix specific QC results aze not included in this report. <br />