Laserfiche WebLink
r <br />SILVER • <br />Method 272.2 (Atomic Absorption, furnace technique) <br />Optimum Concentration Range: 1-25 ug/1 <br />Detection Limit: 0.2 ug/1 <br />STOREY NO. Total 01077 <br />Dissolved 01075 <br />Suspended 01076 <br />`~\ <br />~. <br />Preparation of Standard Solution <br />1. Stock Solution: Prepare as described under "direct aspiration method". <br />2. Prepare dilutions of the stock solution to be used as calibration standards at the time of <br />analysis. These solutions are also to be used for "standard additions". <br />3. The calibration standard should be diluted to contain 0.5%a (v/v) HNO~. <br />Sample Preservation <br />1. For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods <br />section of this manual. <br />Sample Preparation • <br />1. Prepare as described under "direct aspiration method". Sample solutions for analysis <br />should contain 0.5% (v/v) HN03. <br />Instrument Pazameters (General) <br />1. Drying Time and Temp: 30sec-125°C. <br />2. Ashing Time and Temp: 30sec--000°C. <br />3. Atomizing Time and Temp: 10 sec-2700°C. <br />4. Purge Gas Atmosphere: Argon <br />5. Wavelength: 328.1 nm <br />6. Other operating parameters should be set as specified by the particulaz instrument <br />manufacturer. <br />Analysis Procedure <br />1. For the analysis procedure and the calculation, see "Furnace Procedure" pats 9.3 of the <br />Atomic Absorption Methods section of this manual. <br />Approved for NPDES and SD~VA <br />Issued 1978 <br />I <br />272.2-1 <br />