Laserfiche WebLink
middle of the analytical range, a Continuing Calibration Verification (CCV) <br />standard with concentrations neaz the middle of the analytical range but <br />different than those in the ICV, and a readback of the highest calibration <br />standazd. <br />These solutions provide verification that the calibration equations aze <br />functioning properly throughout the analytical range of the instrument. During <br />sample analysis dilutions aze made for analytes found at concentrations above <br />the highest calibration standazd. No results aze taken from extrapolations <br />beyond the highest standard. <br />6. All standards and solutions are NIST traceable and were used within their <br />recommended shelf life. <br />7. The samples were prepared and analyzed within the established hold times. <br />All in house quality control procedures were followed, as described below. <br />8. General quality control procedures. <br />^ A filter (method) blank and laboratory control sample were filtered, preserved <br />and prepazed at the same time as the samples. There were not more than 20 <br />samples associated with each filtered blank and laboratory control sample. The <br />laboratory control sample was only prepared for the selenium analysis. <br />^ The prepazation (method) blank associated with each batch was below the <br />practical quantitation ]unit for each requested analyte. <br />^ The laboratory control sample associated with the selenium batch was within <br />the acceptance limits. <br />^ All initial and continuing calibration blanks associated with this analytical batch <br />were below the practical quantitation limits for the requested analytes. <br />^ All initial and continuing calibration verifications associated with this analytical <br />batch were within the acceptance criteria for the requested analyzes. This <br />indicates a valid calibration and stable instrument conditions. <br />^ The interference check samples associated with Method 200.8 were analyzed. <br />9. Matrix specific quality control procedures. <br />Sample 0508184-1 was designated as the quality control sample for this analysis. <br />Similarity of matrix and therefore relevance of the QC results should not be <br />automatically inferred for any sample other than the native sample selected for QC. <br />^ A matrix spike and matrix spike duplicate were prepared and analyzed with this <br />batch. All acceptance criteria for accuracy were met. <br />^ A matrix spike duplicate was prepazed and analyzed with this batch. All <br />acceptance criteria for precision were met. <br />^ A serial dilution was analyzed with the ICP-MS batch. All acceptance criteria <br />were met. <br />