Paragon Analytics <br />{NORGANICS CASE NARRATIVE <br />CEMEX, Inc. <br />Order Number - 0601125 <br />1. This report consists of 1 water sample. <br />2. The sample was received cool and intact on 01/20106. <br />3. The sample was prepared for analysis based on Methods for the Chemical <br />Analysis of Waters and Wastes (MCAWW), May 1994 procedures. <br />4. The sample was analyzed following MCAWW procedures for the following <br />methods: <br />Anal e Method SOP # <br />pH 150.1 1126 Rev 15 <br />Specific conductance 120.1 1128 Rev 8 <br />5. All standards and solufions were used within their recommended shelf life. <br />6. The sample was prepared and analyzed within the established hold time for each <br />analysis. <br />All in house quality control procedures were followed, as described below. <br />7. General quality control procedures. <br />^ All initial and continuing calibration verifications (ICV/CCV) associated with <br />each analytical batch were within the acceptance criteria for the requested <br />analytes. This indicates a valid calibration and stable instrument conditions. <br />8. Matrix specific quality control procedures. <br />Sample 060 1 1 2 5-2 was designated as the quality control sample for each analysis. <br />Similarity of matrix and therefore relevance of the QC results should not be <br />automatically inferred for any sample other than the native sample selected for <br />QC. <br />^ A sample duplicate was prepared and analyzed with each batch. All guidance <br />criteria for precision were met. <br />