selected as a quality control (QC) sample, matrix specific QC results for the ICP <br />and ICPTrace analyses are not included in this report. <br />^ A matrix spike and matrix spike duplicate were digested and analyzed with <br />each batch specified above. All acceptance criteria for accuracy were met. <br />^ Matrix spike recoveries could not be evaluated for the Following analyses: <br />Anal e <br />Selenium <br />Thallium <br />Sample ID <br />0407037-1 <br />0407037-1 <br />The concentrations of these analytes in the native sample were greater than <br />four times the concentration of matrix spike added during the digestion. <br />When sample concentration is that much greater than the spike added, spike <br />recoveries may not be accurate. The laboratory control sample indicates that <br />the digestion and analysis were in control. <br />A sample duplicate and matrix spike duplicate were digested and analyzed <br />with each batch specified above. All acceptance criteria for precision were <br />met. <br />^ A serial dilution was analyzed with the ICP-MS batch. All acceptance criteria <br />were met. <br />10. It is a standazd PA practice that samples for ICP-MS are analyzed at a dilution. <br />The data contained in the following report have been reviewed and approved by the <br />personnel listed below. In addition, Pazagon Analytics certifies that the analyses reported <br />her 'n aze true, complete and correct within the limits of the methods employed. <br />a <br />D Patrick Date <br />Data Reporting Specialist <br />~,~ M ~ Z ~ <br />Reviewer's Initials Date <br />000 C4 <br />