ACZ Laboratories, Inc. • • - - • - • <br /> 2773 Downhill Drive Steamboat Springs, CO 80487 (800)334-5493 - • • <br /> ACZ Project ID: L65443 <br /> METHOD:ACZ ID WORKNUM PARAMETER QUAL DESCRIPTION <br /> Trichloroethene M8260C/D GUMS RA Relative Percent Difference(RPD)was not used for data <br /> validation because the concentration of the duplicated <br /> sample is too low for accurate evaluation(<10x MDL). <br /> Trichlorofluoromethane M8260C/D GUMS M1 Matrix spike recovery was high,the recovery of the <br /> associated control sample(LCS or LFB)was acceptable. <br /> M8260C/D GUMS RA Relative Percent Difference(RPD)was not used for data <br /> validation because the concentration of the duplicated <br /> sample is too low for accurate evaluation(<10x MDL). <br /> Vinyl Acetate M8260C/D GC/MS RA Relative Percent Difference(RPD)was not used for data <br /> validation because the concentration of the duplicated <br /> sample is too low for accurate evaluation(<10x MDL). <br /> M8260C/D GUMS VB CCV recovery was outside of acceptance limits.See Case <br /> Narrative. <br /> Vinyl Chloride M8260C/D GC/MS M1 Matrix spike recovery was high,the recovery of the <br /> associated control sample(LCS or LFB)was acceptable. <br /> M8260C/D GUMS RA Relative Percent Difference(RPD)was not used for data <br /> validation because the concentration of the duplicated <br /> sample is too low for accurate evaluation(<10x MDL). <br /> REPAD.15.06.05.01 <br /> L65443-2105101543 Page 33 of 37 <br />