Laserfiche WebLink
CRDL Requirements <br /> All CRDL standard(s) met the referenced advisory control limits. <br /> ICSA/ICSAB Statement <br /> All interference check samples (ICSA and ICSAB) associated with this SDG met the <br /> established acceptance criteria. <br /> Continuing Calibration Blank (CCB) Requirements <br /> All continuing calibration blanks (CCB)bracketing this batch met the established <br /> acceptance criteria. <br /> Continuing Calibration Verification (CCV) Requirements <br /> All continuing calibration verifications (CCV)bracketing this SDG met the acceptance <br /> criteria. <br /> Quality Control (QC) Information <br /> Method Blank (MB) Statement <br /> The MBs analyzed with this SDG met the acceptance criteria. <br /> Laboratory Control Sample (LCS) Recovery <br /> The LCS spike recoveries met the acceptance limits. <br /> Quality Control (QC) Sample Statement <br /> The following samples were selected as the quality control (QC) samples for this SDG: <br /> 316524001 (Sunday 1)-ICP, 316515001 (MW-23_12052012)-ICP and CVAA and <br /> 316518004 (MPL 04-1212-1)-ICP-MS. <br /> Matrix Spike (MS) Recovery Statement <br /> The percent recoveries (%R) obtained from the MS analyses are evaluated when the <br /> sample concentration is less than four times (4X) the spike concentration added. All <br /> applicable elements met the acceptance criteria. <br /> Duplicate Relative Percent Difference (RPD) Statement <br /> The RPD obtained from the designated sample duplicate (DUP) is evaluated based on <br /> acceptance criteria of 20%when the sample is >5X the contract required detection limit <br /> (RL). In cases where either the sample or duplicate value is less than 5X the RL, a control <br /> of+/-RL is used to evaluate the DUP results. All applicable analytes met these <br /> requirements. <br /> Serial Dilution % Difference Statement <br /> The serial dilution is used to assess matrix suppression or enhancement. Raw element <br /> concentrations that are 25X the IDL/MDL for CVAA, 50X the IDL/MDL for ICP, and <br /> 100X the IDL/MDL for ICP-MS analyses are applicable for serial dilution assessment. <br /> All applicable analytes met the acceptance criteria of less than 10% difference (%D). <br /> Page 21 of 86 <br />