Laserfiche WebLink
Standards Information <br /> Standard solutions for these analysis are NIST traceable or verified with a NIST traceable standard and used <br /> before the expiration dates. <br /> Sample Geometry <br /> All counting sources were prepared in the same geometry as the calibration standards. <br /> Quality Control(QC)Information: <br /> Blank Information <br /> The blank volume is representative of the sample volume in this batch. <br /> Designated QC <br /> The following sample was used for QC: 316524001 (Sunday 1). <br /> QC Information <br /> All of the QC samples met the required acceptance limits. <br /> Technical Information: <br /> Holding Time <br /> All sample procedures for this sample set were performed within the required holding time. <br /> Sample Re-prep/Re-analysis <br /> None of the samples in this sample set required reprep or reanalysis. <br /> Chemical Recoveries <br /> All chemical recoveries meet the required acceptance limits for this sample set. <br /> Miscellaneous Information: <br /> Data Exception(DER)Documentation <br /> Data exception reports are generated to document any procedural anomalies that may deviate from referenced <br /> SOP or contractual documents.A data exception report(DER)was not generated for this SDG. <br /> Additional Comments <br /> The matrix spike and matrix spike duplicate, 1202797819(Sunday 1)and 1202797820(Sunday 1),aliquots were <br /> reduced to conserve sample volume. <br /> Qualifier Information <br /> Manual qualifiers were not required. <br /> Method/Analysis Information <br /> Product: Gross Alpha/Beta in Drinking Water EPA 900.0 <br /> Analytical Method: EPA 900.0 <br /> Analytical Batch Number: 1274046 <br /> Page 72 of 86 <br />